We use cookies to ensure that we give you the best experience on our website. If you click 'Accept all cookies' we'll assume that you are happy to receive all cookies and you won't see this message again. If you click 'Reject all non-essential cookies' only necessary cookies providing core functionality such as security, network management, and accessibility will be enabled. Click 'Find out more' for information on how to change your cookie settings.

OBJECTIVES: To assess the safety and feasibility of high-intensity focused ultrasound (HIFU) ablation of liver tumours and to determine whether post-operative MRI correlates with intra-operative imaging. METHODS: 31 patients were recruited into two ethically approved clinical trials (median age 64; mean BMI 26 kg m(-2)). Patients with liver tumours (primary or metastatic) underwent a single HIFU treatment monitored using intra-operative B-mode ultrasound. Follow-up consisted of radiology and histology (surgical trial) or radiology alone (radiology trial). Radiological follow-up was digital subtraction contrast-enhanced MRI. RESULTS: Treatment according to protocol was possible in 30 of 31 patients. One treatment was abandoned because of equipment failure. Transient pain and superficial skin burns were seen in 81% (25/31) and 39% (12/31) of patients, respectively. One moderate skin burn occurred. One patient died prior to radiological follow-up. Radiological evidence of ablation was seen in 93% (27/29) of patients. Ablation accuracy was good in 89% (24/27) of patients. In three patients the zone of ablation lay ≤2 mm outside the tumour. The median cross-sectional area (CSA) of the zone of ablation was 5.0 and 5.1 cm(2) using intra-operative and post-operative imaging, respectively. The mean MRI:B-mode CSA ratio was 1.57 [95% confidence interval (CI)=0.57-2.71]. There was positive correlation between MRI and B-mode CSA (Spearman's r=0.48; 95% CI 0.11-0.73; p=0.011) and the slope of linear regression was significantly non-zero (1.23; 95% CI=0.68-1.77; p<0.0001). CONCLUSIONS: HIFU ablation of liver tumours is safe and feasible. HIFU treatment is accurate, and intra-operative assessment of treatment provides an accurate measure of the zone of ablation and correlates well with MRI follow-up.

Original publication

DOI

10.1259/bjr/56737365

Type

Journal article

Journal

Br J Radiol

Publication Date

10/2012

Volume

85

Pages

1363 - 1370

Keywords

Aged, Aged, 80 and over, Feasibility Studies, Female, High-Intensity Focused Ultrasound Ablation, Humans, Intraoperative Care, Liver Neoplasms, Magnetic Resonance Imaging, Male, Middle Aged, Postoperative Care, Prospective Studies, Treatment Outcome